doseBadge Reader
Each Complete Kit is supplied with a Reader
unit. One Reader can be used to control as many doseBadges as you
need.

The Reader is used to carry out the following functions:
- Reset and Calibrate the doseBadge
- Start the doseBadge running (after fitting
it to the worker's shoulder)
- Stop the doseBadge at the end of the day
- Read and Display the noise
measurement results
- Download to computer
| |
|
Calibrating the Dosemeter
The Reader has a built in Calibrator, operating at 1 kHz,
114 dB.
To calibrate the dosemeter, simply push it into the calibration
cavity and press the Cal button on the Reader.
The calibration process is automatic and the results are recorded,
along with any automatic adjustments that are made. |
 |
Configuring the doseBadge
The Reader unit can also be used to configure the doseBadge settings.
The doseBadge can be configured to meet the ISO Q=3 (for European
regulations) and OSHA Q=5 (for the American standards).
The Exchange Rate, Criterion Level, Criterion Time, Threshold
and Slow Time Weighting can be set individually in order to meet
other standards such as those published by MSHA, AICHE and ACGIH.
See the Noise
Regulations page for more information.
As the doseBadge has Dual Channels, it can have Channel 1 set
to OSHA, MSHA, AICHE, ACGIH or similar, with Channel 2 fixed to
measure with the common ISO 3dB settings. |
|