Noise Measurements Made
The main noise measurements taken by the doseBadge depend
on the settings and the regulations being followed. To satisfy
the European Noise at Work regulations, the following parameters
are recorded:
- Lep,d - the worker's daily exposure to noise
- LAeq - the "A" weighted Leq, or average sound level
- LCPeak - the maximum "C" weighted Peak
- The LAeq every minute and the maximum C weighted Peak
every minute
For European noise at work dosimetry, these are all the
parameters that are needed and the doseBadge can be used
with its default settings. However, if you ever have to measure
for other standards, such as the American OSHA standards,
the doseBadge can be configured to measure the TWA, LAvg,
Dose % and others, with exchange rates of 3 dB, 4 dB or 5
dB. Different Criterion Times and Criterion Levels can also
be set.
The doseBadge has dual channels, and can
have Channel 1 configured to have different exchange (doubling)
rate, time response, threshold, criterion level and criterion
time. Channel 2 is fixed to the common ISO 3dB settings.
This means that you can work to regulations such as OSHA
but still record the ISO parameters too. |